Are you a researcher, scientist, or laboratory professional looking for a comprehensive guide to the Narishige PC-10 manual? Look no further! In this article, we'll take a deep dive into the world of precision instruments and explore the features, benefits, and applications of the Narishige PC-10. Whether you're a seasoned user or new to the world of microscopy, this article will provide you with a thorough understanding of the Narishige PC-10 manual and help you unlock the full potential of this remarkable instrument.
| Problem | Probable Cause | Remedy | | :--- | :--- | :--- | | Even if power is on, "HEATER LEVEL" does not display. | AC Power Cord is disconnected; voltage setting is incorrect; fuse is blown. | Plug cord in firmly; make correct voltage setting; check and replace fuse. | | Heater value is the same, but the pipette is different. | Heating Element is deteriorated. | Increase the heater value to compensate for loss of heat generation. | | Pipette looks distorted and off-center. | Glass capillary is not centered in the Heating Element. | Be sure the glass capillary is placed right at the center of the Heating Element. | | Heating function is inconsistent. | Heater Unit is affected by wind from AC, room temperature, and humidity. | Please place the instrument to avoid those factors as much as possible. | | The heating function is inconsistent and coils are touching. | The coils of Heating Element are in contact. | Use a wrench to separate them, or replace it with a new unit of PC-10H. |
The Narishige PC-10 is a dual-stage vertical glass micropipette puller widely recognized for its reliability in producing consistent, high-quality microelectrodes and micropipettes. This comprehensive manual and operating guide provides step-by-step instructions for installation, configuration, and advanced parameter tuning to achieve optimal results for various electrophysiology and microinjection applications. 1. System Overview and Technical Specifications narishige pc10 manual new
: Use a high first-stage temperature to initiate melting and a lower second-stage temperature to refine the tip resistance. 5. Maintenance and Troubleshooting
(or equivalent 1.0mm O.D.) glass capillary through the heater coil. Are you a researcher, scientist, or laboratory professional
Secure the glass capillary tubes using knurled thumb screws. The lower slider moves vertically via gravity.
Securely plug in the AC cord; check and replace fuses in the inlet. Internal controller malfunction. Contact Narishige Service. Uneven Pulls Dirty filament or misaligned glass. Whether you're a seasoned user or new to
Understanding the physical layout is key to navigating the manual 1.2.1.
Carefully remove the upper and lower pipettes vertically to avoid snapping the fragile tips.